OCT Drive Characterization Report

Spec: OCT catheter drive — example acceptance spec
Overall result: PASS

Engineering bench characterization of drive mechanics only. No clinical, diagnostic, or medical-efficacy claim is made.

Acceptance limits

Parameter Measured Limit Check Result
NURD index 0.7213 % 5 % <= limit PASS
RPM mean error 0.0005845 % (abs) 2 % (abs) RPM error
Rotational jitter (period std) 0.1283 % 2 % <= limit PASS
Pullback speed error 0.5 % (abs) 3 % (abs) speed error

Supporting measurements

NURD

  • Overall NURD index: 0.7213 %
  • Per-rotation NURD (mean / max): 0.7004 % / 1.018 %
  • Dominant NURD harmonic: 1x (ratio 0.004899)
  • Revolutions analyzed: 100

Rotational stability

  • RPM mean: 6000
  • RPM std: 7.701
  • Period std / peak-to-peak: 0.1283 % / 0.6384 %
  • Wow / flutter RMS: 0.004598 % / 0.1282 %

Pullback

  • Commanded / measured speed: 20 / 20.1 mm/s
  • Speed error: 0.5 %
  • Speed CV: 1.14e-10 %; max deviation: 4.757e-10 %
  • Linearity R^2: 1